Vacuum Leak Test for Autoclave:
- To check for leakage in autoclaves, vacuum leak tests are conducted.
- “The Chamber Integrity Test” is another name for the Vacuum Leak test.
- In (Vacuum Leak test) VLT, a vacuum is initially applied within the chamber, followed by a wait for the vacuum to reach a predetermined set point, a holding period during which the vacuum is maintained, and finally, vacuum breakage.
- Check the results afterward to see if the autoclave leakage was within the acceptable range.
- According to the Health Technical Memorandum 2010 (HTM 2010) standards, VLT should be done once per week. (Full form of HTM is Health Technical Memorandum)
- However, it is advisable to carry out VLT on a daily basis because this is a very important test that determines the chamber’s integrity.
- There is a possibility that non-sterile air will enter the autoclave chamber if there is any leakage there or if the allowed amount of leakage is exceeded.
- When a vacuum break occurs, air enters the chamber through a 0.22-micron filter that is situated on the autoclaves’ unloading side.
- As sterile air passes through a 0.22-micron filter before entering the chamber, there is no danger of contamination.
- However, if air entered the chamber through any other chamber connector than the filter, non-sterile air would enter and potentially contaminate the sterilized load.
Acceptance criteria for Vacuum Leak Test (VLT):
- As per Health Technical Memorandum 2010 (HTM 2010) standards Vacuum Leak Test (VLT) should be not more than (NMT) 1.3 mbar per minute.
- However, some autoclave printouts specify the acceptance requirement as 0.013 bar without specifying whether the vacuum leakage limit is per minute or per 10 minutes. The technician can become confused as a result.
- Here is the clarity:.
For example:
- Vacuum at hold time start: A
- Vacuum at hold time end: B
- Actual leakage: A-B= Z per minute
- The result for vacuum leakage per 10 minutes is given by Z X 10 (Result for Leakage for 10 min.)
- As per HTM 2010, the Acceptance criteria is NMT 1.3 mbar per minute
- Some autoclave printouts suggest the results as 0.013 bar
- 1 bar is equal to 1000 mbar or 0.001 bar per min. or 0.010 bar per 10 min.
- so, 1.3 mbar per min. or 0.0013 bar per min. or 0.013 bar per 10 minutes.
Read More:
- Sterilization in Pharmaceuticals
- Autoclaving
- Difference between a Class B and Class N Autoclaves
- Bowie Dick Test
- Vacuum leak test in Autoclaves
- Types of Autoclaves
- LAL test
- LAL test Procedure
- TLC & HPTLC
- GxP in Pharmaceuticals
- Types of chromatography
- CIP and SIP in Pharmaceuticals
- Positive control and negative control in Microbiology
- Types of biological indicators for sterilization
- Sterility Assurance Level (SAL)
- D value and Z values in Microbiology
- Vacuum Leak Test (VLT) in Autoclave, Acceptance Criteria and Its Importance In Pharmaceuticals