Calibration of Disintegration Apparatus (DT):
Calibration of DT includes the Sensor calibration, Strokes and Timer calibration. Details are as described below,
A. Calibration of the temperature (Sensor Calibration):
Calibration of the sensor should be performed by an external certified agency or a Highly calibrated Thermometer on below Temperature (°C),
Acceptance criteria:
Set Value | Acceptance Criteria |
30°C | 28-32 °C |
60°C | 58-62°C |
B. Calibration of strokes (up/down cycle):
Calibration of stroke should be performed by setting 1 min and count the strokes. Perform the calibration Three time and take a average. Acceptance limit are as shown below,
Limits: 29-32 strokes per minute.
C. Calibration of timer:
Calibration of the Timer of DT shall be measured by means of a calibrated stopwatch. Perform the calibration Three time for each set time. Limit is described below,
Acceptance criteria:
Sr. No. | Set Time (in Min.) | Limit (For Stop Watch) (min:sec) |
1 | 1 | 00:59 – 00:61 |
2 | 5 | 04:55 – 05:05 |
3 | 10 | 09:50 – 10:10 |
4 | 15 | 14:45 – 15:15 |
Calibration Frequency: The disintegration Apparatus should be calibrated yearly basis.
Cleaning of Disintegration Apparatus (DT):
- Remove water from the water bath periodically and clean the instrument and bath with clean and dry cloth.
- Use mild detergents for cleaning of the disintegration assembly
Read More:
- Alcohol Dose Dumping
- Methods to Compare Dissolution Profiles
- 21 CFR 11
- Documentation Practice (GDP) & ALCOA++
- Peak Purity
More Topic on:
- Quality Assurance
- Analytical Development or Quality Control
- Formulation Development
- Health Topic
- For Vitamins & Minerals:
Reference:
ICH: https://www.ich.org/
USFDA: https://www.fda.gov/regulatory-information/search-fda-guidance-documents
EMA: https://www.ema.europa.eu/en/human-regulatory/research-development/scientific-guidelines