Calibration of Friability test apparatus

How do you calibrate the Friability Apparatus or Friability Tester or Friabilitor?

Calibration Procedure of friability test apparatus as follow,

  • Calibration will be performed on Time mode and Count mode of the friability Tester or friabilitor.

A. Calibration by Time mode:      

  • Set the time by selecting the Time mode and setting the time to 1 minute.
  • Then start the apparatus and check the No. of Revolutions in 1 minute.
  • Repeat the test by setting the time for 2 minutes, 4 minutes and check the No. of Revolutions in 2 minutes, 4 minutes and record the result.
  • Each test is to be performed in triplicate.

 Acceptance criteria:

  • 24 to 26 rotations in 1 minute.
  • 48 to 52 rotations in 2 minutes.
  • 96 to 104 rotations in 4 minutes.

B. Calibration by Count mode:      

  • Set the count by selecting count mode and set the count to 25 revolutions.
  • Start the apparatus and check the time with the help of a stopwatch complete 25
  • Repeat the test by setting the count to 50 revolutions, 100 revolutions, and checking the time with the help of a stopwatch to complete 50 revolutions, 100 revolutions, and record the result.
  • Each test is to be performed in triplicate.

Acceptance criteria:

  • 0:57 00 to 1:03 00 minutes for 25 rotations.
  • 1:54 00 to 2:06 00 minutes for 50 rotations.
  • 3:48 00 to 4:12 00 minutes for 100 rotations.

Calibration of Friability test apparatus

Calibration Frequency:

It should be calibrated on a six-monthly basis.