Calibration of DT includes the Sensor calibration, Strokes and Timer calibration. Details are as described below,
Calibration of the sensor should be performed by an external certified agency or a Highly calibrated Thermometer on below Temperature (°C),
Acceptance criteria:
| Set Value | Acceptance Criteria |
| 30°C | 28-32 °C |
| 60°C | 58-62°C |
Calibration of stroke should be performed by setting 1 min and count the strokes. Perform the calibration Three time and take a average. Acceptance limit are as shown below,
Limits: 29-32 strokes per minute.
Calibration of the Timer of DT shall be measured by means of a calibrated stopwatch. Perform the calibration Three time for each set time. Limit is described below,
Acceptance criteria:
| Sr. No. | Set Time (in Min.) | Limit (For Stop Watch) (min:sec) |
| 1 | 1 | 00:59 – 00:61 |
| 2 | 5 | 04:55 – 05:05 |
| 3 | 10 | 09:50 – 10:10 |
| 4 | 15 | 14:45 – 15:15 |
Calibration Frequency: The disintegration Apparatus should be calibrated yearly basis.